HEMATOLOGY ANALYZER

URIT BH-6180 - 5 DIFF HEMATOLOGY ANALYZER

URIT BH-6180

5 PART DIFF

  • 5-PART-DIFF Hematology Analyzer
  • Flow Cytometry + Nucleic Acid Fluorescence Staining Technology
  • One-stop capillary blood testing
  • Outperforming ‘6-Part-Diff’ hematology analyzer
  • 44 reportable parameters and 61 research parameters
  • New PLT-L solution ensure the accuracy of low PLT count

SPECIFICATIONS

Parameters37 Reportable parameters (whole blood): WBC, Lym(#,%), Mon(#,%), Neu(#,%), Bas(#,%),
Eos(#,%), IMG(#,%), RBC, HGB, HGT, MCV, MCH, MCHC, RDW-CV, RDW-SD, NRBC(#,%),
PLT, MPV, PDW, PCT, P-LCR, P-LCC, RET(#,%), IRF, HFR, MFR, LFR, IPF, RHE
With 2 histograms, 5 scatter grams, 3 3D stereograms
50 Research parameters (whole blood): HPC(#,%), HFC(#,%), RBC-O, PLT-O, PLT-I,
WBC-O, WBC-D, TNC-D, IME(#,%), H-NR%, L-NR%, NLR, PLR, WBC-N, TNC-N, lnR#,
lnR%o, Micro(#,%), Macro(#,%), MRV, PDW-SD, RPI, H-IPF, IPF#, Neu-X, Neu-Y, Neu-Z,
Lym-X, Lym-Y, Lym-Z, Mon-X, Mon-Y, Mon-Z, SRBC, LRBC, SMCV, LMCV, MCH(r),
HDW, MPC, MPM, HYPER%, HYPO%, FRC(#,%); With 4 scatter grams
7 Reportable parameters (body fluid): WBC-BF, RBC-BF, TC-BF#, MN(#,%), PMN(#,%)
11 Research parameters (body fluid): Eos-BF(#,%), Neu-BF(#,%), HF-BF(#,%), RBC-BF*,
LY-BF(#,%), MO-BF(#,%)
Body fluid with RBC histogram and DIFF scatter gram
Channel and PrincipleDIFF/WNR/RET/WPC: Flow Cytometry+Fluorescence Nucleic Acid Staining technology
RBC(PLT): Sheath Impedance
HGB: Colorimetric(The upgrated Cyanide-free reagent)
ThroughtputCBC 110T/H, CD 110T/H, CDR 65T/H
ReagentsDiluents(DS, DR), lyses(LN, LD, LP, LH), Fluorescent dyes(FR, FN, FD, FP)
Sampling modePre-diluent 20 μL, Whole blood≤80 μL, Capillary blood≤36 μL, Body fluid≤85 μL
Sampling deviceAuto loader with 50 (10*5 racks) samples storage capacity
Data storageUp to 1,500,000 sample results with grams
Analysis modeCBC, CD, CDH, CDR, CDRH, CR, RET, CR/PL T -L, CDR/PL T-L
LinearityWBC: 0~500x109/L, PL T: 0~5000x109/L
DisplayIntegrated 10.4-inch touch screen with PC display (23.8-inch color TFT display)
HIS & LISSupport, with HL7 protocol
PowerAC 100-240V, 50/60Hz, 400VA
Dimension & Weight570mm(L) x 775mm(W) x 725mm(H), 80kg